기본 정보
연구 분야
발행물
구성원
article|
인용수 0
·2017
A study on a pedicle-screw-based reduction method for artificially reduced artifacts
Hyunju Kim, Hae-Kag Lee, Jae-Hwan Cho
Journal of the Korean Physical Society
초록

The purpose of this study is a quantitative analysis of the degree of the reduction of the artifacts that are induced by pedicle screws through the application of the recently developed iterative metallic artifact reduction (I MAR) software. Screw-type implants that are composed of 4.5 g/cm3 titanium (Ti) with an approximate average computed tomography (CT) value of 6500 Hounsfield units (HUs) that are used for the treatment of spinal diseases were placed in paraffin, a tissueequivalent material, and then dried. After the insertion, the scanning conditions were fixed as 120 kVp and 250 mA using multidetector computed tomography (MDCT) (Enlarge, Siemens, Germany). The slice thickness and the increment were set at the fields of view (FOVs) of 3 mm and 120 mm, respectively; the pitch is 0.8; the rotation time is 1 s; and the I MAR software was applied to the raw data of the acquired images to compare the CT-value changes of the posterior images. When the I MAR software was applied to animal vertebrae, it was possible to reduce the 65.7% image loss of the black-hole-effect image through the application of the I MAR software. When the I MAR image loss (%) was compared with the white-streak-effect image, the high-intensity image type with the white-streak effect could be reduced by 91.34% through the application of the I MAR software. In conclusion, a metal artifact that is due to a high-density material can be reduced more effectively when the I MAR algorithm is applied compared with that from the application of the conventional MAR algorithm. The I MAR can provide information on the various tissues that form around the artifact and the reduced metal structures, which can be helpful for radiologists and clinicians in their determination of an accurate diagnosis.

키워드
Hounsfield scaleStreakMaterials scienceReduction (mathematics)Artifact (error)SoftwareNuclear medicineImaging phantomBiomedical engineeringComputed tomography
타입
article
IF / 인용수
- / 0
게재 연도
2017